Color Knowledge
  • Lubricant Generated Insoluble Color Bodies Test Method


  • Standard Test Method for
    Measurement of Lubricant Generated Insoluble Color
    Bodies in in-service Turbine Oils using Membrane Patch
    Colorimetry

    This standard is issued under the fixed designation D7843;the number immediately following the indicates the year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicateas the year of last reapproval. A superscript epsilon (ε) indicates an editorial change since the last revision or reapproval.

    1. Scope*
      1.1 This test method extracts insoluble contaminants from a
    samnnple of in-service turbine oil onto a patch and the color of the membrane patch is analyzed by a spectrophotometer. The
    results are reported as a ΔE value, within the CIE LAB scale.                                                                                                                                                            
      1.2 This test method is not appropniate for turbine oils with dyes.
      1.3 The values stated in SI units are to be regarded as
    standard. No other units of measurement are included in this
    standard.
      1.4 This standard does not purport to address all of the
    safery concems, if any, associated with its use. It is the
    responsibility of the user of this standard to establish appropriate safer and health practices and determine the applica
    bility of regulator limitations prior to use.

    2. Apparatus
      2.1 Vanations of apparatus, particularly with respect to filter membranes and vacuum pump setting, can affect the test results significantly.
      2.2 When the user of this test method uses an altemate
    membrane filter. It is incumbent upon them to establish that the alternate filter will give equal results.
      2.3 Regared Apparatus:
      2.3.1 Membrane Filter, 47 mm nitro-cellulose,0.45µm.
      2.3.2 Forceps, smooth-trp.
      2.3.3 Borosilicate Filter Holder.
      2.3.4 Borosilicate Filtering Flask.
      2.3.5 Wash Bottle equipped with 0.22 µm membranes.
      2.3.6 Vacuum Source, capable of maintaining a vacuum of
    71kPa土5kPa.
      2.3.7 Graduated cylinder, 150 ml to 200 ml.
      2.3.8 Beaker, 100 ml to 250 ml.
      2.3.9 Petri dish.
      2.3.10 Spectrophotometer, with capabilities of analyzing a standard 15 mm target with a 0?/45? measuring geometry, 10? Observer, 10 nm spectral intervals minimum resolution, the visible spectral range of 400 nm to 700 nm and CIELAB
    measuring indices.